Nano-Micro Letters | |
Self-Powered Implantable Skin-Like Glucometer for Real-Time Detection of Blood Glucose Level In Vivo | |
Huiling Gao1  Shuai Wang2  Wanglinhan Zhang2  Lili Xing2  Xinyu Xue2  Linlin Zhang2  Wenyan Yang2  | |
[1] College of Life and Health Sciences, Northeastern University;College of Sciences, Northeastern University; | |
关键词: Diabetes; Biosensor; Electronic-skin; Self-powered; Glucose detection; Implantable electronics; | |
DOI : 10.1007/s40820-017-0185-x | |
来源: DOAJ |
【 摘 要 】
Abstract Implantable bioelectronics for analyzing physiological biomarkers has recently been recognized as a promising technique in medical treatment or diagnostics. In this study, we developed a self-powered implantable skin-like glucometer for real-time detection of blood glucose level in vivo. Based on the piezo-enzymatic-reaction coupling effect of GOx@ZnO nanowire, the device under an applied deformation can actively output piezoelectric signal containing the glucose-detecting information. No external electricity power source or battery is needed for this device, and the outputting piezoelectric voltage acts as both the biosensing signal and electricity power. A practical application of the skin-like glucometer implanted in mouse body for detecting blood glucose level has been simply demonstrated. These results provide a new technique path for diabetes prophylaxis and treatment.
【 授权许可】
Unknown