期刊论文详细信息
Measurement: Sensors
Development and measurement of a 365 NM UV LED irradiance source
Muyao Ji1  Fangsheng Lin2  Leibing Shi3  Biyong Huang3  Ming Xia3  Tiecheng Li3  Liping Wang3  Weihai Cheng3 
[1] Shanghai Institute of Measurement and Testing Technology, Shanghai, PR China;Corresponding author.;Shanghai Institute of Measurement and Testing Technology, Shanghai, PR China;
关键词: UV LED;    Irradiance source;    Integrating sphere;    Calibration;   
DOI  :  
来源: DOAJ
【 摘 要 】

UV LED has been widely used in various fields such as medical disinfection and biochemical detection, especially in UV curing. Compared with traditional high-voltage mercury lamp, UV LED shows large differences both in the spectral power distribution and irradiance, which may cause up to 20% error in calibration. A 365 nm UV LED source has been developed and measured, which is made of UV LEDs, temperature control module, integrating sphere, circuit and monitoring module. The experiment results indicate that the irradiance of about 884 mW/cm2 meets the actual requirements of the curing industry, and the stability of 1.0% (15 min) and uniformity of 1.6% are in compliance with the calibration regulation.

【 授权许可】

Unknown   

  文献评价指标  