| Earth and Space Science | |
| Calibration of the Heat Flow and Physical Properties Package (HP 3) for the InSight Mars Mission | |
| T. Spohn1  M. Grott1  J. Knollenberg1  M. Scharringhausen2  T. Wippermann2  T.L. Hudson3  W.B. Banerdt3  S.E. Smrekar3  C. Krause4  | |
| [1] Institute of Planetary Research German Aerospace Center (DLR) Berlin Germany;Institute of Space Systems German Aerospace Center (DLR) Bremen Germany;Jet Propulsion Laboratory California Institute of Technology Pasadena CA USA;Space Operations and Astronaut Training German Aerospace Center (DLR) Cologne Germany; | |
| 关键词: Mars; Heat Flow; Geopysics; InSight; | |
| DOI : 10.1029/2019EA000670 | |
| 来源: DOAJ | |
【 摘 要 】
Abstract The (HP 3) of the InSight Mars mission is an instrument package to measure the Martian geothermal heat flow. The instrument's platinum resistance temperature sensors have been calibrated between −75 and +55 °C, and the absolute temperature uncertainty of the calibration is 10 mK. Temperature differences can be measured with an uncertainty of better than 5 mK. HP 3 determines regolith thermal conductivity by using the HP 3 mole as a modified line heat source, and thermal conductivity can be measured with an uncertainty of better than 4%. Tilt sensors used to determine the attitude of the mole during penetration measure tilt with an uncertainty of better than 0.25°.
【 授权许可】
Unknown