Quality assurance of simultaneous treatment of two targets in pelvic region planned with single isocenter using three dimensional conformal radiotherapy (3DCRT) technique | |
Vadhiraja Bejadi Manjunath1  Sridhar Chinthamani2  Arun Kumar Elavunkal Sukumaran3  Prakash Saxena3  Challapalli Srinivas3  Sourjya Banerjee3  Suman Kumar Putha3  Dinesh Pai Kasturi3  | |
[1] Department Radiation Oncology, Manipal Hospital, Bangalore;Department of Radiation Oncology, Father Muller Oncology Center, Mangalore;Department of Radiotherapy & Oncology, Kasturba Medical College Hospital, Mangalore; | |
关键词: Quality Assurance; Two Targets; Single Isocenter; Conformal Radiotherapy; | |
DOI : 10.14319/ijcto.33.15 | |
来源: DOAJ |
【 摘 要 】
Purpose: The purpose of this study was to conduct quality assurance of a three dimensional conformal radiotherapy (3DCRT) of two targets in pelvis region planned with single isocenter technique.
Methods: A treatment plan was generated with two identical water phantoms with ionization chamber (IC) sleeves (IC-1 & IC-2), simulated as if targets are in pelvis region, simultaneously irradiated with single isocenter technique with a dose prescription of 300 cGy for point dose verification. A two dimensional ion chamber array detector was used for fluence verification.
Results: Calculated minimum, mean and maximum dose (in cGy) for IC-1 & IC-2 were 295, 303 and 307 as per dose volume histogram. The global dose maximum was found to be 307.4 cGy. Measured point doses to both lesions were within ±2.5% of the computed dose. A pass percentage of 97% was obtained with the set of criteria 3 mm distance to agreement and 3% dose difference for fluence verification.
Conclusion: Treatment execution of two targets simultaneously with single isocenter can reduce positional errors and delivery time.