Revista Brasileira de Ortopedia | |
Biomechanical analysis of a double fixation method for tendon graft in porcine tibia - using an interference screw plus staple | |
关键词: Anterior cruciate ligament; Tibia; Orthopedic fixation devices; Biomechanical phenomena; Tendons; | |
DOI : 10.1016/j.rboe.2018.07.003 | |
来源: DOAJ |
【 摘 要 】
ABSTRACTObjective: The aim of the study was to compare the mechanical behavior of interference screw tibial fixation vs. screw-plus-staple tibial fixation in an animal model.Methods: Thirty-six pieces of swine knee specimens were selected and divided into two groups: Group 1, tibial fixation with interference screw (n = 17), and Group 2, fixation with interference screw and staple (n = 19). The models were submitted to a single cycle of tension testing. The following variables were measured: graft cross-sectional area, failure point on 10 mm (F10), yield load (Fy), and stiffness.Results: The mean values of graft cross-sectional area, F10, Fy, and stiffness did not present significant differences between the groups.Conclusion: The addition of a second staple-type ligament fixation device, complementing the interference screw, did not increase the mechanical safety of the system.
【 授权许可】
Unknown