期刊论文详细信息
Sensors
Motion Artifact Reduction in Wearable Photoplethysmography Based on Multi-Channel Sensors with Multiple Wavelengths
Minseong Kim1  Jongshill Lee1  InYoung Kim1  Hoon-Ki Park2 
[1] Department of Biomedical Engineering, Hanyang University, Seoul 04763, Korea;Department of Family Medicine, Hanyang University, Seoul 04763, Korea;
关键词: photoplethysmography;    motion artifact;    independent component analysis;    multi-wavelength;   
DOI  :  10.3390/s20051493
来源: DOAJ
【 摘 要 】

Photoplethysmography (PPG) is an easy and convenient method by which to measure heart rate (HR). However, PPG signals that optically measure volumetric changes in blood are not robust to motion artifacts. In this paper, we develop a PPG measuring system based on multi-channel sensors with multiple wavelengths and propose a motion artifact reduction algorithm using independent component analysis (ICA). We also propose a truncated singular value decomposition for 12-channel PPG signals, which contain direction and depth information measured using the developed multi-channel PPG measurement system. The performance of the proposed method is evaluated against the R-peaks of an electrocardiogram in terms of sensitivity (Se), positive predictive value (PPV), and failed detection rate (FDR). The experimental results show that Se, PPV, and FDR were 99%, 99.55%, and 0.45% for walking, 96.28%, 99.24%, and 0.77% for fast walking, and 82.49%, 99.83%, and 0.17% for running, respectively. The evaluation shows that the proposed method is effective in reducing errors in HR estimation from PPG signals with motion artifacts in intensive motion situations such as fast walking and running.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次