期刊论文详细信息
Sensors & Transducers
Aluminum-Selective Electrode Based on (1E,2E)-N1,N2- dihydroxy )-N1,N2- bis (4-hydroxy phenyl) Oxalimidamide as a Neutral Carrier
Esfandyar Baghdar1  Fekri Lari F.1  Aghaie H.2  Aghaie M.3  Ali Kakanejadifard4 
[1] Ph.D.Student, Department of Chemistry,Science and Research Branch, Islamic Azad University, Tehran, Iran;Department of Chemistry,Science and Research Branch, Islamic Azad University, Tehran, Iran,;Faculty of Chemistry, North Tehran Branch, Islamic Azad University, Tehran, Iran;Organic Chemistry, No30, Amiri Alley, Motahary Street, Khorramabad, Iran;
关键词: Ionophore;    Additive;    Limit of detection;    Ion-selective electrode;    Nernstian slope;   
DOI  :  
来源: DOAJ
【 摘 要 】

A synthesized,(1E,2E)-N1, N2- dihydroxy-N1, N2-bis (4-hydroxyphenyl) oxalimidamide (DBO) , has been used as an ionophore in order to preparation of a new ion-selective electrode. An optimized membrane was prepared by mixing of 70 mg PVC powdered, 2 mg of considered ionophore, 100 mg of plasticizer and 2 mg of additive. The sensor exhibits reversible potential responses with an almost Nernstian slope for Al3+ ions in the concentration range from 1.0 ´ 10-6 M to 1.30´10-1 M. The limit of detection (LOD) was 2.50 ´ 10-7 M The electrode’s response is independent of pH in the range of 3.2-7.8. The response time of the electrode is about 15 s and can be used for 50 days without any divergence in potential. The practical application of the electrode was tested by using it as indicator electrode to determine the end point in the potentiometric titration of Al3+ with EDTA solution.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:1次