Commercially Fabricated Printed Circuit Board Sensing Electrodes for Biomarker Electrochemical Detection: The Importance of Electrode Surface Characteristics in Sensor Performance | |
Despina Moschou1  Gorachand Dutta1  Anna Regoutz2  | |
[1] Centre for Biosensors, Bioelectronics and Biodevices (C3Bio), Department of Electronic & Electrical Engineering, University of Bath, Bath BA2 7AY, UK; | |
[2] Department of Materials, Imperial College London, Exhibition Road, London SW7 2AZ, UK; | |
关键词: electrode pretreatment; electrochemical biosensor; printed-circuit-board; glucose detection; chronoamperometry; | |
DOI : 10.3390/proceedings2130741 | |
来源: DOAJ |
【 摘 要 】
Here we report the first PCB-implemented electrochemical glucose biosensor using
covalently immobilized glucose oxidase (GOx) on the commercially fabricated PCB electrode
surface, taking particular care on the electrode surface characteristics and their effect on sensor
performance. Based on the results, this assay exhibits a highly linear response from 500 μM to 20
mM (R = 0.9961) and a lower limit of detection of 500 μM.