Sensors | 卷:22 |
Low-Cost, Real-Time Polymerase Chain Reaction System for Point-of-Care Medical Diagnosis | |
Yvonne Sun1  Chengkun Liu2  Vamsy P. Chodavarapu2  Tchamie Kadja2  | |
[1] Department of Biology, University of Dayton, 300 College Park, Dayton, OH 45469, USA; | |
[2] Department of Electrical and Computer Engineering, University of Dayton, 300 College Park, Dayton, OH 45469, USA; | |
关键词: polymerase chain reaction (PCR); fluorescence sensing; low-cost PCR; proportional–integral–derivative (PID) controller; point-of-care diagnostics; COVID-19; | |
DOI : 10.3390/s22062320 | |
来源: DOAJ |
【 摘 要 】
Global health crises due to the prevailing Coronavirus Disease 2019 (COVID-19) pandemic have placed significant strain on health care facilities such as hospitals and clinics around the world. Further, foodborne and waterborne diseases are not only spreading faster, but also appear to be emerging more rapidly than ever before and are able to circumvent conventional control measures. The Polymerase Chain Reaction (PCR) system is a well-known diagnostic tool for many applications in medical diagnostics, environmental monitoring, and food and water quality assessment. Here, we describe the design, development, and testing of a portable, low-cost, and real-time PCR system that can be used in emergency health crises and resource-poor situations. The described PCR system incorporates real-time reaction monitoring using fluorescence as an alternative to gel electrophoresis for reaction analysis, further decreasing the need of multiple reagents, reducing sample testing cost, and reducing sample analysis time. The bill of materials cost of the described system is approximately $340. The described PCR system utilizes a novel progressive selective proportional–integral–derivative controller that helps in reducing sample analysis time. In addition, the system employs a novel primer-based approach to quantify the initial target amplicon concentration, making it well-suited for food and water quality assessment. The developed PCR system performed DNA amplification at a level and speed comparable to larger and more expensive commercial table-top systems. The fluorescence detection sensitivity was also tested to be at the same level as commercially available multi-mode optical readers, thus making the PCR system an attractive solution for medical point-of-care and food and water quality assessment.
【 授权许可】
Unknown