Inductive powering of subcutaneous stimulators: key parameters and their impact on the design methodology | |
Adrien Debelle1  Carmen Godfraind2  Antoine Nonclercq2  Vicente Acuña3  Pascal Doguet3  Laurent Lonys3  | |
[1] Université Libre de Bruxelles; | |
[2] Synergia Medical, Mont-Saint-Guibert; | |
关键词: Implanted neurostimulators; Inductive powering; Design methodology; Efficiency optimisation; | |
DOI : 10.4081/ejtm.2016.6040 | |
来源: DOAJ |
【 摘 要 】
Inductive powering of implantable medical devices involves numerous factors acting on the system efficiency and safety in adversarial ways. This paper lightens up their role and identifies a procedure enabling the system design. The latter enables the problem to be decoupled into four principal steps: the frequency choice, the magnetic link optimization, the secondary circuit and then finally the primary circuit designs. The methodology has been tested for the powering system of a device requirering a power of 300mW and implanted at a distance of 15 to 30mm from the outside power source. It allowed the identification of the most critical parameters. A satisfying efficiency of 34% was reached at 21mm and tend to validate the proposed design procedure.