期刊论文详细信息
Computer Assisted Surgery 卷:27
Optimal definitions for computing HKA angle in caos: an in-vitro comparison study
Bhushan Borotikar1  Guillaume Dardenne1  Hoel Letissier1  Eric Stindel1  Ahmed Zemirline2 
[1] CHU Brest, Brest, France;
[2] Centre Hospitalier Privé, France;
关键词: Knee center;    frontal plane;    HKA;    CAOS;   
DOI  :  10.1080/24699322.2022.2066023
来源: DOAJ
【 摘 要 】

The goal of this study was to assess and compare the precision and accuracy of nine and seven methods usually used in Computer Assisted Orthopedic Surgery (CAOS) to estimate respectively the Knee Center (KC) and the Frontal Plane (FP) for the determination of the HKA angle (HKAA). An in-vitro experiment has been realized on thirteen cadaveric lower limbs. A CAOS software application was developed and allowed the computation of the HKAA according to these nine KC and seven FP methods. The precision and the accuracy of the HKAA measurements were measured. The HKAA precision was highest when the FP is determined using the helical method. The HKAA accuracy was highest using the helical approach to determine the FP and either the notch or the tibial spines to determine the KC. This study shows that the helical approach to determine the FP and either the notch or the middle of tibia spines are the combinations that provide both a good enough accuracy and precision to estimate the HKA.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次