| Chromatography | 卷:2 |
| Field-Amplified Sample Injection-Micellar Electrokinetic Chromatography for the Determination of Benzophenones in Food Simulants | |
| Oscar Núñez1  Cristina Félez1  Anna Molet1  | |
| [1] Department of Analytical Chemistry, University of Barcelona, Martí i Franquès 1-11, E08028 Barcelona, Spain; | |
| 关键词: field-amplified sample injection; micellar electrokinetic chromatography; benzophenones; migration studies; food simulants; | |
| DOI : 10.3390/chromatography2030452 | |
| 来源: DOAJ | |
【 摘 要 】
A field-amplified sample injection-micellar electrokinetic chromatography (FASI-MEKC) method for the determination of 14 benzophenones (BPs) in a food simulant used in migration studies of food packaging materials was developed, allowing almost baseline separation in less than 21 min. The use of a 10 mM sodium dodecyl sulfate (SDS) solution as sample matrix was mandatory to achieve FASI enhancement of the analyzed BPs. A 21- to 784-fold sensitivity enhancement was achieved with FASI-MEKC, obtaining limits of detection down to 5.1–68.4 µg/L, with acceptable run-to-run precisions (RSD values lower than 22.3%) and accuracy (relative errors lower than 21.0%). Method performance was evaluated by quantifying BPs in the food simulant spiked at 500 µg/L (bellow the established specific migration limit for BP (600 µg/L) by EU legislation). For a 95% confidence level, no statistical differences were observed between found and spiked concentrations (probability at the confidence level, p value, of 0.55), showing that the proposed FASI-MEKC method is suitable for the analysis of BPs in food packaging migration studies at the levels established by EU legislation.
【 授权许可】
Unknown