期刊论文详细信息
Chemosensors 卷:9
An Aptamer-Array-Based Sample-to-Answer Biosensor for Ochratoxin A Detection via Fluorescence Resonance Energy Transfer
Min Fang1  Bowen Yang1  Chenling Tang1  Min Xiao1  Yaxi Li1  Wenlian Wang1  Shu Zhang1  Tianfu Wu1  Yu Wang1  Gongjun Tan1  Zhenfei Peng2  Yongning Li3 
[1] Department of Biomedical Engineering, University of Houston, Houston, TX 77204, USA;
[2] Department of Medical Technology, Fujian Health College, Fuzhou 350101, China;
[3] Department of Pharmacy, Fujian Health College, Fuzhou 350101, China;
关键词: aptamer;    ochratoxin A;    FRET;    on-chip assay;   
DOI  :  10.3390/chemosensors9110309
来源: DOAJ
【 摘 要 】

Food toxins are a hidden threat that can cause cancer and tremendously impact human health. Therefore, the detection of food toxins in a timely manner with high sensitivity is of paramount importance for public health and food safety. However, the current detection methods are relatively time-consuming and not practical for field tests. In the present work, we developed a novel aptamer-chip-based sample-to-answer biosensor (ACSB) for ochratoxin A (OTA) detection via fluorescence resonance energy transfer (FRET). In this system, a cyanine 3 (Cy3)-labeled OTA-specific biotinylated aptamer was immobilized on an epoxy-coated chip via streptavidin-biotin binding. A complementary DNA strand to OTA aptamer at the 3′-end was labeled with a black hole quencher 2 (BHQ2) to quench Cy3 fluorescence when in proximity. In the presence of OTA, the Cy3-labeled OTA aptamer bound specifically to OTA and led to the physical separation of Cy3 and BHQ2, which resulted in an increase of fluorescence signal. The limit of detection (LOD) of this ACSB for OTA was 0.005 ng/mL with a linearity range of 0.01–10 ng/mL. The cross-reactivity of ACSB against other mycotoxins, ochratoxin B (OTB), aflatoxin B1 (AFB1), zearalenone (ZEA), or deoxynilvalenol (DON), was less than 0.01%. In addition, this system could accurately detect OTA in rice samples spiked with OTA, and the mean recovery rate of the spiked-in OTA reached 91%, with a coefficient of variation (CV) of 8.57–9.89%. Collectively, the ACSB may represent a rapid, accurate, and easy-to-use platform for OTA detection with high sensitivity and specificity.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次