期刊论文详细信息
Iranian Journal of Microbiology 卷:7
Rapid fluorometric quantification of bacterial cells using Redsafe nucleic acid stain
Mahdi Aminian1  Ehsan Khalili2  Roya Solhi2  Vahid Hosseini2 
[1] Department of Clinical Biochemistry, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran AND Recombinant Vaccine Research Center, Tehran University of Medical Sciences, Tehran, Iran.;
[2] Department of Clinical Biochemistry, School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.;
关键词: Fluorometric assay;    Redsafe;    MTT assay;    E.coli;   
DOI  :  
来源: DOAJ
【 摘 要 】

Background and Objectives: Numerous procedures in biology and medicine require the counting of cells. Direct enumer- ation of Colony Forming Units (CFUs) is time-consuming and dreary accurate cell counting on plates with high numbers of CFUs is error prone. In this study we report a new indirect cell counting method that was developed based on the use of Redsafe fluorometric assay. The usefulness of Redsafe, a nucleic acid stain, in liquid medium is based on the binding of the fluorescent dye to DNA.Materials and Methods: Redsafe fluorometric assay was evaluated in comparison with MTT colorimetric assay as a colou- rimetric assay for enumeration of bacterial cells.Results: Obtained results showed that fluorometric assay threshold for LB grown E. coli is 6×104  CFU/ml. Redsafe fluo- rescent assay can be used as a rapid and inexpensive method for bacterial enumeration and quantification with increased sensitivity.Conclusion: The sensitivity of the Redsafe fluorometric assay for detection and enumeration of bacterial cells was 2-log-unit more than that was observed for the MTT assay.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次