期刊论文详细信息
The European Physical Journal C | |
Krypton assay in xenon at the ppq level using a gas chromatographic system and mass spectrometer | |
Hardy Simgen1  Sebastian Lindemann1  | |
[1] Max-Planck-Institut für Kernphysik, Saupfercheckweg 1, 69117 , Heidelberg, Germany; | |
关键词: Xenon; Systematic Uncertainty; Krypton; Procedure Blank; Cold Finger; | |
DOI : 10.1140/epjc/s10052-014-2746-1 | |
来源: Springer | |
【 摘 要 】
We have developed a new method to measure krypton traces in xenon at unprecedented low concentrations. This is a mandatory task for many near-future low-background particle physics detectors. Our system separates krypton from xenon using cryogenic gas chromatography. The amount of krypton is then quantified using a mass spectrometer. We demonstrate that the system has achieved a detection limit of 8 ppq (parts per quadrillion) and present results of distilled xenon with krypton concentrations below 1 ppt.
【 授权许可】
CC BY
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202112165959017ZK.pdf | 386KB | download |