Clinical evaluation of an authorized medical equipment based on high performance liquid chromatography for measurement of serum voriconazole concentration
Kaho Matsuyama1  Shota Uchino1  Mai Horikawa1  Kayo Kurogi1  Naoya Matsumoto1  Hitomi Arakaki1  Kou Yonemaru1  Taiki Katsume1  Tomomi Katanoda2  Kazutaka Oda2  Yuki Narita2  Koji Iwamura2  Hirofumi Jono3  Hideyuki Saito3 
[1] Department of Clinical Pharmaceutical Sciences, Graduate School of Pharmaceutical Sciences, Kumamoto University, 1-1-1, Honjo, Chuo-ku, Kumamoto, Japan;Department of Pharmacy, Kumamoto University Hospital, 1-1-1, Honjo, Chuo-ku, Kumamoto, Japan;Department of Pharmacy, Kumamoto University Hospital, 1-1-1, Honjo, Chuo-ku, Kumamoto, Japan;Department of Clinical Pharmaceutical Sciences, Graduate School of Pharmaceutical Sciences, Kumamoto University, 1-1-1, Honjo, Chuo-ku, Kumamoto, Japan;
关键词: Voriconazole;    High performance liquid chromatography;    Therapeutic drug monitoring;   
DOI  :  10.1186/s40780-021-00225-8
来源: Springer
PDF
【 摘 要 】

BackgroundTherapeutic drug monitoring for voriconazole is recommended for its optimum pharmacotherapy. Although the feedback of the measurement result of serum voriconazole concentration by outsourcing needs a certain time (days within a 1 week), there was no medical equipment for the measurement available in clinical practice. Recently, a medical equipment based on high performance liquid chromatography, named LM1010, has been developed and authorized for clinical use. In this study, to validate the clinical performance of LM1010, we compared the measured serum voriconazole concentrations by LM1010 with those by outsourcing measurement using liquid chromatography-tandem mass spectrometry.MethodsWe conducted the observational study approved by the institutional review board of Kumamoto University Hospital (No. 1786). Residual serum samples harvested for therapeutic drug monitoring were separated. Measured concentrations by LM1010 by the standard filter method (needs serum volume of > 400 μL) or the dilute method (needs serum volume of 150 μL) were compared with those by outsourcing, respectively. Acceptable measurement error range of 0.72–1.33 was considered. There were 69 serum samples, where the 35 or 34 samples were employed for evaluation of the standard filter method or the dilute method, respectively.ResultsThe measured concentration using the standard filter method/outsourcing was 2.22/2.10 μg/mL as the median, 1.57–3.40/1.53–3.62 as the interquartile range, < 0.2–10.76/< 0.2–11.46 μg/mL as the range, while those using the dilute method/outsourcing was 2.36/2.29 μg/mL as the median, 1.08–2.94/1.03–3.06 as the interquartile range, 0.24–10.00/< 0.2–10.85 μg/mL as the range. The regression line for the standard filter method or the dilute method were y = 0.935x + 0.154 or y = 0.933x + 0.162, respectively. The standard filter method or the dilute method showed 11.4% samples (4/35, 95%CI 3.2–26.7%) or 8.8% samples (3/34, 95%CI 1.9–23.7%) out of the acceptable measurement error range, respectively.ConclusionMeasurement of serum voriconazole concentration by LM1010 can be acceptable in clinical TDM practice.

【 授权许可】

CC BY   

【 预 览 】
附件列表
Files Size Format View
RO202112046723116ZK.pdf 620KB PDF download
  文献评价指标  
  下载次数:0次 浏览次数:2次