Dual-energy computed tomography of the neck—optimizing tube current settings and radiation dose using a 3D-printed patient phantom
article
Torsten Diekhoff1  Michael Scheel1  Wiebke Kress1  Bernd Hamm1  Paul Jahnke2 
[1] Department of Radiology, Charité - Universitätsmedizin Berlin, Campus Mitte, Humboldt-Universität zu Berlin, Freie Universität Berlin;Berlin Institute of Health (BIH)
关键词: Phantoms;    imaging;    printing;    three-dimensional;    tomography;    X-ray computed;    radiation exposure;   
DOI  :  10.21037/qims-20-854
学科分类:外科医学
来源: AME Publications
PDF
【 摘 要 】

Background: Dual-energy computed tomography (DECT) is increasingly used in studies and clinical practice. However, the best protocol is controversially discussed and whether it exhibits more radiation exposure compared to conventional protocols. Thus, the purpose of the study was to determine optimal tube current settings for DECT in a 3D-printed anthropomorphic phantom of the neck. Methods: A 3D-printed iodinated ink based phantom of a contrast enhanced CT of the neck was imaged. Six dual-energy multi-detector computed tomography scans were performed with six different tube currents (80 kVp: 30–400 mAs; 135 kVp: 5–160 mAs). 120 virtual blended images (VBIs) and 66 virtual monochromatic images (VMIs) were reconstructed and 12 regions of interest (bilaterally: common carotid arteries, subcutaneous soft tissue, mandibular bone, sternocleidomastoid muscle, submandibular gland, and mid-image: vertebral body of C2 and pharyngeal space) in six consecutive slices resulting in 96 measurements per scan were performed. Hounsfield units and signal- and contrast-to-noise ratio were compared to single-energy computed tomography as standard of reference. Results: VBIs overestimated the Hounsfield units (P<0.0001). Optimal dual-energy scanning parameters resulted in 120% (100 kVe: 51.2 vs. 61.7 and 65.2, for signal and contrast-to-noise ratio, respectively; 120 kVe: 60.8 vs. 72.1 vs. 128.3) of the radiation exposure with about 80% of the signal/contrast-to-noise ratio of the corresponding single-energy images. However, optimal weighting of tube currents for both voltages depended on the desired reconstruction. Conclusions: Dual-energy protocols apply an estimated 120% of the single-energy radiation exposure and result in approximately 80% of the image quality. Tube current settings should be adapted to the desired information.

【 授权许可】

All Rights reserved   

【 预 览 】
附件列表
Files Size Format View
RO202108210002510ZK.pdf 1577KB PDF download
  文献评价指标  
  下载次数:4次 浏览次数:0次