期刊论文详细信息
Journal of Veterinary Internal Medicine
Evaluation of an Electrochemical Point‐of‐Care Meter for Measuring Glucose Concentration in Blood from Periparturient Dairy Cattle
A.A. Megahed2  M.W.H. Hiew2  J.R. Townsend2  J.B. Messick1 
[1] Department of Comparative Pathobiology, College of Veterinary Medicine, Purdue University, West Lafayette, IN;Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Purdue University, West Lafayette, IN
关键词: hyperglycemia;    hyperketonemia;    hypoglycemia;    ketosis;   
DOI  :  10.1111/jvim.13608
来源: Wiley
PDF
【 摘 要 】

Abstract

Background

The Precision Xtra® meter is a promising low cost electrochemical point-of-care unit for measuring blood glucose concentration ([gluc]) in cattle blood. The meter uses an algorithm that assumes the intra-erythrocyte [gluc] equals the plasma [gluc] on a molal basis, and that the hematocrit is similar in humans and cattle.

Objectives

The primary objective was to determine the accuracy of the meter for measuring plasma [gluc] in dairy cattle. Secondary objectives were to characterize the influence of hematocrit and sample temperature on the measured value for [gluc].

Animals

A total of 106 periparturient Holstein-Friesian cattle.

Methods

Blood and plasma samples (1,109) were obtained and Deming regression and Bland–Altman plots were used to determine the accuracy of the meter against the reference method (plasma hexokinase assay). Multivariable regression and linear regression were used to determine the effect of hematocrit and sample temperature on the plasma [gluc] measured by the meter.

Results

Intra-erythrocyte [gluc] was 18% of plasma [gluc] on a molar basis. Sample temperature had a significant linear effect on plasma [gluc] as measured by the meter for 3/5 plasma samples when measured [gluc] > 160 mg/dL.

Conclusions and clinical importance

The meter utilizes an algorithm that is optimized for human blood and is inaccurate when applied to bovine blood. Until a cattle-specific algorithm is developed, we recommend using plasma as the analyte instead of blood and calculating plasma [gluc] using the equation: [gluc] = 0.66 × [gluc]p-meter + 15, where [gluc]p-meter is the value reported by the meter. If blood is measured, then we recommend using the equation: [gluc] = 0.90 × [gluc]b-meter + 15.

【 授权许可】

CC BY-NC   
Copyright © 2015 The Authors. Journal of Veterinary Internal Medicine published by Wiley Periodicals, Inc. on behalf of the American College of Veterinary Internal Medicine.

Creative Commons Attribution-NonCommercial License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.

【 预 览 】
附件列表
Files Size Format View
RO202107150014422ZK.pdf 3727KB PDF download
  文献评价指标  