期刊论文详细信息
Journal of Medical Radiation Sciences
An imaging evaluation of the simultaneously integrated boost breast radiotherapy technique
Jessica Turley B MedRadSci (RT)1 
[1] Department of Radiation Oncology, Royal Prince Alfred Hospital, Camperdown, New South Wales, Australia
关键词: Breast radiotherapy;    imaging;    simultaneous integrated boost;    treatment margins;    treatment verification;   
DOI  :  10.1002/jmrs.117
来源: Wiley
PDF
【 摘 要 】

Abstract

Introduction

To evaluate in-field megavoltage (MV) imaging of simultaneously integrated boost (SIB) breast fields to determine its feasibility in treatment verification for the SIB breast radiotherapy technique, and to assess whether the current-imaging protocol and treatment margins are sufficient.

Methods

For nine patients undergoing SIB breast radiotherapy, in-field MV images of the SIB fields were acquired on days that regular treatment verification imaging was performed. The in-field images were matched offline according to the scar wire on digitally reconstructed radiographs. The offline image correction results were then applied to a margin recipe formula to calculate safe margins that account for random and systematic uncertainties in the position of the boost volume when an offline correction protocol has been applied.

Results

After offline assessment of the acquired images, 96% were within the tolerance set in the current department-imaging protocol. Retrospectively performing the maximum position deviations on the Eclipse treatment planning system demonstrated that the clinical target volume (CTV) boost received a minimum dose difference of 0.4% and a maximum dose difference of 1.4% less than planned. Furthermore, applying our results to the Van Herk margin formula to ensure that 90% of patients receive 95% of the prescribed dose, the calculated CTV margins were comparable to the current departmental procedure used.

Conclusion

Based on the in-field boost images acquired and the feasible application of these results to the margin formula the current CTV-planning target volume margins used are appropriate for the accurate treatment of the SIB boost volume without additional imaging.

【 授权许可】

CC BY-NC-ND   
© 2015 The Authors. Journal of Medical Radiation Sciences published by Wiley Publishing Asia Pty Ltd on behalf of Australian Institute of Radiography and New Zealand Institute of Medical Radiation Technology.

Creative Commons Attribution-NonCommercial-NoDerivs License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non-commercial and no modifications or adaptations are made.

【 预 览 】
附件列表
Files Size Format View
RO202107150014276ZK.pdf 1320KB PDF download
  文献评价指标  
  下载次数:4次 浏览次数:17次