期刊论文详细信息
Frontiers in Cardiovascular Medicine
Proof of Concept: Measuring Aortic Annulus Resistance by Means of Pressure-Volume Curves During Balloon Inflation to Guide Transcatheter Aortic Valve Implantation
Timothée Noterdaeme1  Felix Vogt1  Nikolaus Marx1  Jonas Gesenhues2  Steffen Massberg3 
[1] Department of Cardiology & Intensive Care Medicine, University Hospital Aachen, RWTH Aachen University, Aachen, Germany;Institute of Automatic Control, RWTH Aachen University, Aachen, Germany;Munich University Center, Ludwig-Maximilians University, and German Centre for Cardiovascular Research (DZHK), Partner Site Munich Heart Alliance, Munich, Germany;
关键词: aortic valve stenosis;    transcatheter aortic valve implantation;    pressure volume curve;    atrioventicular block;    paravalvular leak;    translational research;   
DOI  :  10.3389/fcvm.2021.665029
来源: Frontiers
PDF
【 摘 要 】

This study assessed the basic working principle to measure aortic annulus resistance during balloon inflation for transcatheter aortic valve implantation (TAVI), by acquisition of pressure-volume curve for a guided semi-automatic implantation. A modular bench-system was used which allows the incremental inflation of valvuloplasty balloons by means of a stepper-motor driven linear axis with simultaneous recording of the pressure changes inside the system. Different porcine aortic xenografts were assessed by use of a non-compliant valvuloplasty balloon. In a second step transcatheter aortic stents were implanted inside target sized xenografts. The recorded pressure volume-curves showed that the system can accurately differentiate between different xenografts and assess the quality of the tissue rendering real-time analysis of pressure-volume curves during balloon-inflation possible, which has the potential to optimize the implantation procedure by direct adaptation to the patient specific anatomy and characteristics. Further investigations and development are warranted.

【 授权许可】

CC BY   

【 预 览 】
附件列表
Files Size Format View
RO202107132367548ZK.pdf 750KB PDF download
  文献评价指标  
  下载次数:20次 浏览次数:2次