BMC Anesthesiology
Rupture of an epidural filter connector during bolus administration of local anesthetic: a case report
Aaron R. Muncey1  Jamie P. Hoffman1  Daniel A. Nahrwold1  Nasrin N. Aldawoodi1  Raymond M. Evans1 
[1] H. Lee Moffitt Cancer Center & Research Institute, University of South Florida Morsani College of Medicine, 12902 USF Magnolia Dr, 33612, Tampa, FL, USA;
关键词: Epidural catheter;    Filter connector;    Equipment;    Local anesthetic;    Pressure;    Safety;    Injury;    Infection;   
DOI  :  10.1186/s12871-021-01372-z
来源: Springer
PDF
【 摘 要 】

BackgroundEpidural catheters are routinely placed for many surgical procedures and to treat various pain conditions. Known complications arising from epidural catheter equipment malfunction include epidural pump failure, epidural catheter shearing, epidural catheter connector failure, epidural filter connector cracking, and loss-of-resistance syringe malfunction. Practitioners need to be aware of these potentially dangerous complications and take measures to mitigate the chances of causing significant patient harm. We report on the complete breakage of an epidural filter connector during epidural bolus administration of local anesthetic by hand with a syringe.Case presentationA B. Braun Perifix® epidural catheter was placed in a 73-year-old male scheduled for radical prostatectomy. During the operation, a continuous infusion of local anesthetic was administered through the epidural catheter in addition to general endotracheal anesthesia. At the conclusion of surgery and after extubation, the patient endorsed incisional pain. The epidural filter connector broke in half as a bolus of local anesthetic was administered by hand with a syringe. The local anesthetic sprayed widely throughout the room as the fragmented epidural filter connector became a projectile object that recoiled and struck the patient.ConclusionsThis incident placed the patient and surrounding healthcare providers at substantial risk for injury and infection from the fractured epidural filter connector becoming a projectile object and from the local anesthetic spray. The most plausible cause of this event was from a large amount of pressure being applied to the filter connector. This may have occurred by excessive force being applied by hand to the syringe, by the presence of a clogged filter, or by the catheter being kinked or blocked proximal to the filter. Being aware of this deleterious complication and potentially modifying existing epidural bolus techniques, such as using smaller syringes with less applied force and checking all epidural components vigilantly prior to and during bolus administration, can help anesthesia providers deliver the safest possible care to patients with epidural catheters.

【 授权许可】

CC BY   

【 预 览 】
附件列表
Files Size Format View
RO202107078863021ZK.pdf 383KB PDF download
  文献评价指标  
  下载次数:2次 浏览次数:2次