3D Printing in Medicine
Low cost 3D printed clamps for external fixator for developing countries: a biomechanical study
Jose Nauaki Shiozawa1  Arthur Erdman2  Cara Piazza2  Felix J. Landaeta2 
[1] Centro de Salud B La Troncal, Instituto Ecuatoriano de Seguridad Social, Quito, Ecuador;Earl E. Bakken Medical Devices Center University of Minnesota-Twin Cities, 420 Delaware Street SE, MMC 95, G217 Mayo Building, 55455, Minneapolis, MN, USA;
关键词: Linear external fixation;    Biomechanical properties;    Low cost;    3D print;    Third world;    Developing countries;    ASTM F1541;   
DOI  :  10.1186/s41205-020-00084-3
来源: Springer
PDF
【 摘 要 】

BackgroundExternal fixation is a mainstream limb reconstruction technique, most often used after a traumatic injury. Due to the high rates of trauma in developing countries, external fixation devices are often utilized for immediate fracture stabilization and soft tissue repair. Proper external fixation treatment too often still fails to be adopted in these regions due to the high cost and trauma complexity. A novel, inexpensive, unilateral fixator was constructed using 3D printed clamps and other readily available supporting components. ASTM standard F1541 tests were used to assess the biomechanical properties of this novel external fixator.MethodsApplicable sections of ASTM standard F1541 were used to determine the biomechanical properties of the novel external fixator. 3D printed clamps modeled using SolidWorks and printed with chopped carbon fibers using a fuse deposition modeling (FDM) based 3D printer by Markforged (Boston, MA) were used. This study included 3 different testing configurations: axial compression, anterior-posterior (AP) bending, and medial-lateral (ML) bending. Using the novel unilateral fixator with 3D printed clamps previously sterilized by autoclave, an input load was applied at a rate of 20 N/s, starting at 0 N via a hydraulic MTS tester Model 359. Force and deformation data were collected at a sampling rate of 30 Hz. There was a load limit of 750 N, or until there was a maximum vertical deformation of 6 mm. Also, 4 key dimensions of the 3D printed clamps were measured pre and post autoclave: diameter, width, height and length.ResultsThe novel external fixator had axial compression, AP and ML bending rigidities of 246.12 N/mm (σ = 8.87 N/mm), 35.98 N/mm (σ = 2.11 N/mm) and 39.60 N/mm (σ =2.60 N/mm), respectively. The 3D printed clamps shrunk unproportionally due to the autoclaving process, with the diameter, width, height and length dimensions shrinking by 2.6%, 0.2%, 1.7% and 0.3%, respectively.ConclusionOverall, the biomechanical properties of the novel fixator with 3D printed clamps assessed in this study were comparable to external fixators that are currently being used in clinical settings. While the biomechanics were comparable, the low cost and readily available components of this design meets the need for low cost external fixators in developing countries that current clinical options could not satisfy. However, further verification and validation routines to determine efficacy and safety must be conducted before this novel fixator can be clinically deployed. Also, the material composition allowed for the clamps to maintain the appropriate shape with minimal dimensional shrinkage that can be accounted for in clamp design.

【 授权许可】

CC BY   

【 预 览 】
附件列表
Files Size Format View
RO202104279942720ZK.pdf 1541KB PDF download
  文献评价指标  
  下载次数:11次 浏览次数:10次