期刊论文详细信息
Química Nova
Potentiometric determination of Nickel (II) ion using 2-hydroxy-1-naphthylidene-N-cyanoacetohydrazone as electroactive material
Harish Kumar Sharma2  Pernita Dogra2  Akhilesh Kumar Gupta2  Fardad Koohyar1 
[1] ,Maharishi Markendeshwer University Engineering College Department of ChemistryAmbala,India
关键词: ion selective electrode;    Schiff's base;    Ni2+;    FIM;    potentiometric titration;   
DOI  :  10.5935/0100-4042.20140198
来源: SciELO
PDF
【 摘 要 】

A potentiometric Nickel sensor was prepared using 2-hydroxy-1-naphthylidene-N-cyanoacetohydrazone as electro-active material and epoxy resin as a binding material. A membrane composed of 40% Schiff's base and 60% epoxy resin exhibited the best performance. The membrane showed excellent response in the concentration range of 0.15 ppm to 0.1 mol L- 1 Ni+2 ions with non-Nernstian slope of 22.0 mV/decade, had a rapid response time (less than 10 s), and can be used for three months without any considerable loss of potential. The sensor was useful within the pH range of 1.3 to 9.6, and was able to discriminate between Ni2+ and a large number of alkaline earth and transition metal ions. The practical utility of the sensor has been demonstrated by using it successfully as an indicator electrode in the potentiometric titration of Ni2+ with EDTA and oxalic acid.

【 授权许可】

CC BY-NC   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202103040064617ZK.pdf 265KB PDF download
  文献评价指标  
  下载次数:2次 浏览次数:1次