期刊论文详细信息
Química Nova
Desenvolvimento de um dispositivo de baixo custo para medidas por quimiluminescência
Eduardo P. Borges2  Elizabeth N. Fernandes2  Fábio R. P. Rocha2  Boaventura F. Reis1 
[1] ,Universidade Federal de São Carlos Departamento de Química São Carlos SP
关键词: flow analysis;    chemiluminescence;    hydrogen peroxide;    photodiode;   
DOI  :  10.1590/S0100-40422002000700021
来源: SciELO
PDF
【 摘 要 】

A simple and low cost device (ca. US$ 150) that comprises two photodiodes fixed in lab-made Perspex flow cell is proposed for chemiluminescence measurements. The characteristics of the device (large observation window and reduced thickness) allow maximizing the amount of the emitted radiation detected. A sensitivity improvement of ca. 50 % was observed by employing two photodiodes for signal measurements. The performance of the device was assessed by the oxidation of luminol by hydrogen peroxide, yielding a linear response within the range of 2.50 to 500 µmol L-1 H2O2. The detection limit was estimated as 0.8 µmol L-1 hydrogen peroxide which is comparable with those obtained by using equipments based on photomultipliers.

【 授权许可】

CC BY-NC   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202103040061141ZK.pdf 256KB PDF download
  文献评价指标  