期刊论文详细信息
Brazilian Dental Journal
Connecting accuracy of interchanged prosthetic abutments to different dental implants using scanning electron microscopy
Piero Rocha Zanardi1  Bruno Costa1  Roberto Chaib Stegun1  Newton Sesma1  Matsuyoshi Mori1  Dalva Cruz Laganá1 
[1] ,University of São Paulo Dental School Department of ProsthodonticsSão Paulo SP ,Brazil
关键词: dental abutments;    dental implants;    dental prosthesis design;    scanning electron microscopy;    prosthesis fitting;   
DOI  :  10.1590/S0103-64402012000500006
来源: SciELO
PDF
【 摘 要 】

The present study evaluated the interchangeability of prosthetic components for external hexagon implants by measuring the precision of the implant/abutment (I/A) interface with scanning electron microscopy. Ten implants for each of three brands (SIN, Conexão, Neodent) were tested with their respective abutments (milled CoCr collar rotational and non-rotational) and another of an alternative manufacturer (Microplant) in randomly arranged I/A combinations. The degree of interchangeability between the various brands of components was defined using the original abutment interface gap with its respective implant as the benchmark dimension. Accordingly, when the result for a given component placed on an implant was equal to or smaller then that gap measured when the original component of the same brand as the implant was positioned, interchangeability was considered valid. Data were compared with the Kruskal-Wallis test at 5% significance level. Some degree of misfit was observed in all specimens. Generally, the non-rotational component was more accurate than its rotational counterpart. The latter samples ranged from 0.6-16.9 µm, with a 4.6 µm median; and the former from 0.3-12.9 µm, with a 3.4 µm median. Specimens with the abutment and fixture from Conexão had larger microgap than the original set for SIN and Neodent (p<0.05). Even though the latter systems had similar results with their respective components, their interchanged abutments did not reproduce the original accuracy. The results suggest that the alternative brand abutment would have compatibility with all systems while the other brands were not completely interchangeable.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130109837ZK.pdf 894KB PDF download
  文献评价指标  
  下载次数:10次 浏览次数:1次