A comparison of amperometric and spectrophotometric methods for the iodide concentration measurement: a tracer in produced water from offshore oil reservoirs
Flávia C. Souza2  Davi Augusto I. Da Silva2  Márcia S. Ribeiro2  Roberto B. Faria2  Maria A. Melo2  Roberta M. M. Toledo1  Eliane D'elia2 
[1] ,Universidade Federal do Rio de Janeiro Instituto de Química Rio de Janeiro RJ ,Brazil
关键词: iodide;    reservoir tracer;    amperometry;    ultramicroelectrode;    spectrophotometry;    produced water;   
DOI  :  10.5935/0103-5053.20130199
来源: SciELO
PDF
【 摘 要 】

Amperometric and spectrophotometric methods were developed for measuring iodide tracer concentration in water produced from offshore oil reservoirs. The amperometric method exhibited a linear response from 1 to 10 mg L-1 with limits of detection and of quantification of 0.2 and 0.6 mg L-1, respectively. The spectrophotometric method also exhibited a linear response from 1 to 10 mg L-1 with limits of detection and of quantification of 1 and 3 mg L-1, respectively. Both methods showed to be accurate, linear, homoscedastic and their recoveries were 101 ± 2 and 100 ± 3%, respectively, while the recoveries for produced reservoir water samples (fortified with iodide) were 97 ± 7 and 100 ± 3%, respectively. High amounts of sulfate interfere in the spectrophotometric method. Both methods are promising, simple and inexpensive alternatives to ion chromatography for the detection of iodide tracer in produced water reservoir samples.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130107181ZK.pdf 566KB PDF download
  文献评价指标  
  下载次数:5次 浏览次数:11次