期刊论文详细信息
Journal of the Brazilian Chemical Society
A flow-batch analyzer for UV-Vis spectrophotometric detection of adulteration in distilled spirits
Elaine C. L. Nascimento2  Mário C. U. Araújo1  Roberto K. H. Galvão1 
[1] ,Universidade Federal Rural de PernambucoSerra Talhada PE ,Brazil
关键词: flow-batch analyzer;    SIMCA;    UV-Vis spectrophotometry;    adulteration of distilled spirits;    screening analysis;   
DOI  :  10.1590/S0103-50532011000600009
来源: SciELO
PDF
【 摘 要 】

An automatic flow-batch analyzer is proposed for detection of adulteration in distilled spirits (whiskey, brandy, cachaça, rum and vodka) using UV-Vis spectrophotometry. The analyzer was employed to dilute the non-adulterated samples and to simulate adulteration with addition of 5 and 10% (v/v) of water, ethanol or methanol. SIMCA (soft independent modelling of class analogies) classification models were built using spectra of non-adulterated and adulterated samples in the region of 235-355 nm. By applying the SIMCA models to a test set, all adulterated and non-adulterated samples were correctly discriminated at a confidence level of 95% with an analytical throughput larger than 120 samples per hour.