Journal of the Brazilian Chemical Society | |
Automatic photometric titration procedure based on multicommutation and flow-batch approaches employing a photometer based on twin LEDs | |
Milton B. Da Silva2  Carla C. Crispino2  Boaventura F. Reis1  | |
[1] ,Universidade Federal de São Carlos Departamento de Química São Carlos SP ,Brazil | |
关键词: multicommuted flow injection analysis; flow-bath analysis; automated photometric titration; fruit juice; LED photometer; | |
DOI : 10.1590/S0103-50532010001000009 | |
来源: SciELO | |
【 摘 要 】
In this work, a reliable and inexpensive photometer based on twin LEDs assembled as a radiation source and as a photodetector is described. The setup including the photometer and flow system module was designed to implement an automated titration procedure employing the multicommuted flow injection analysis (MCFIA) process. The proposed system was able to carry out photometric titration without using analytical curve to achieve the sample concentration. Its usefulness was proven by analyzing vinegar and lemon, orange, pineapple, maracock, and acajou juices. Accuracy was accessed comparing results with those obtained employing a manual procedure and no significant difference was observed at 95% confidence level. Samples presenting different acid concentrations such as lemon juice (1084.73 mmol L-1) and pineapple juice (14.42 mmol L-1) could be titrated using the same titration solution without any modification of the titrating setup.
【 授权许可】
CC BY
All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202005130106309ZK.pdf | 268KB | download |