期刊论文详细信息
Journal of the Brazilian Chemical Society
Arsenic determination in pharmaceutical grade barium sulfate using direct solid sampling electrothermal atomic absorption spectrometry
Rodrigo C. Bolzan2  Diogo P. De Moraes1  Julio Cezar P. De Mattos1  Valderi L. Dressler1  Érico M. De Moraes Flores1 
[1] ,Universidade Regional Integrada do Alto Uruguai e das Missões Departamento de Ciências da Saúde Frederico Westphalen RS ,Brazil
关键词: arsenic determination;    solid sampling;    barium sulfate;    electrothermal atomic absorption spectrometry;    pharmaceutical analysis;   
DOI  :  10.1590/S0103-50532010000400015
来源: SciELO
PDF
【 摘 要 】

A method for arsenic determination in pharmaceutical grade BaSO4 by solid sampling electrothermal atomic absorption spectrometry (SoS-ET AAS) is described. Two background (BG) correction systems, deuterium and Zeeman-effect (Z), were evaluated. Background was decreased by using hydrogen combined with Pd. Accurate and precise results were obtained using Z-SoS-ET AAS. Relative standard deviation using Z-SoS-ET AAS was lower than 9.5%. Limit of detection was 0.005 μg g-1 and characteristic mass was 25.4 pg. Results were also compared with those obtained using the preparation method suggested by European Pharmacopoeia for BaSO4 followed by determination using inductively coupled plasma optical emission spectrometry and inductively coupled plasma mass spectrometry. Results were around 20% lower than results obtained using the proposed method. Some advantages as simplicity, higher sample throughput were found when compared with the official method currently described in pharmacopoeia.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130106160ZK.pdf 431KB PDF download
  文献评价指标  
  下载次数:14次 浏览次数:12次