期刊论文详细信息
Journal of the Brazilian Chemical Society
Calibration techniques and modifiers for the determination of Cd, Pb and Tl in biodiesel as microemulsion by graphite furnace atomic absorption spectrometry
Jessee S. Azevedo Silva2  Eduardo S. Chaves1  Éder J. Dos Santos1  Tatiana D. Saint'pierre1  Vera Lúcia A. Frescura1  Adilson J. Curtius1 
[1] ,Instituto Federal de Educação, Ciência e Tecnologia de Santa CatarinaFlorianópolis SC ,Brazil
关键词: GF AAS;    biodiesel;    microemulsion;    cadmium;    lead;    thallium;    chemical modifier;   
DOI  :  10.1590/S0103-50532010000400006
来源: SciELO
PDF
【 摘 要 】

Different calibration approaches and modifiers were tested for Cd, Pb and Tl determination in biodiesel by graphite furnace atomic absorption spectrometry (GF AAS). Microemulsions were prepared by mixing 2 g of biodiesel, 1 mL of a 10% (v/v) nitric acid aqueous solution and n-propanol to a 10 mL final volume. Pyrolysis temperatures of 600 °C for Cd and Tl with Pd as permanent modifier, and 800 °C for Pb with Ru as permanent modifier were used. Atomization temperatures were also optimized. Calibration solution prepared in aqueous medium did not correct non-spectral interferences, but matrix matching calibration using base oil led to accurate results with recoveries from 80 to 116%. The RSD values were lower than 8% for Cd and Pb and the LOD values were 0.5 ng g-1 for Cd, 6 ng g-1 for Pb and 1 ng g-1 for Tl. The obtained results using the analyte additions method or by matrix matching calibration were in agreement, confirming the accuracy of the proposed procedure. Organometallic standards were not required and different samples were analyzed.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130106151ZK.pdf 416KB PDF download
  文献评价指标  
  下载次数:20次 浏览次数:15次