期刊论文详细信息
Journal of the Brazilian Chemical Society
Development and validation of a multivariate calibration model for determination of dipyrone in oral solutions by near infrared spectroscopy
Marcus H. Ferreira2  Jorge F. F. Gomes2  Marcelo M. Sena1 
[1] ,Indústria Química do Estado de Goiás SAGoiânia GO ,Brazil
关键词: multivariate calibration;    quality control;    PLS;    metamizole;    near infrared spectroscopy;   
DOI  :  10.1590/S0103-50532009000900017
来源: SciELO
PDF
【 摘 要 】

This work developed a new method for determination of dipyrone (DIP) in oral pharmaceutical formulations, through the use of near infrared (NIR) transflectance measurements and multivariate calibration. The studied range varied from 300.0 to 569.2 mg mL-1. The best PLS (partial least squares) model was obtained with two latent variables and the root mean square errors of calibration and prediction were 1.1 and 1.0 mg mL-1, respectively. The proposed method was validated in accordance with ANVISA, the Brazilian regulatory agency, and ICH, being considered selective, linear, precise, accurate and robust. By comparison with the main alternatives, iodimetric titration and HPLC, this method is simpler, non-destructive, does not use reagents or solvents and does not produce chemical waste. Besides, its rapidity is considered the major advantage over the other methods, since only about 50 s were spent per assay.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130106033ZK.pdf 325KB PDF download
  文献评价指标  
  下载次数:3次 浏览次数:7次