期刊论文详细信息
Journal of the Brazilian Chemical Society
Wavelet-Based determination of Cu and Pb in water samples using potentiometric stripping analysis with a Lab-Made potentiostat
Valdomiro L. Martins2  Roberto K. Harrop Galvão1  Mário César U. De Araújo2  Edvaldo Da Nóbrega Gaião1 
[1] ,Universidade Federal da Paraíba Departamento de Química João Pessoa PB ,Brazil
关键词: potentiometric stripping analysis;    wavelet transform;    lead;    copper;    water analysis;   
DOI  :  10.1590/S0103-50532009000800025
来源: SciELO
PDF
【 摘 要 】

This paper demonstrates the possibility of simultaneously determining lead and copper in mineral, river and tap water by potentiometric stripping analysis using a low-cost lab-made potentiostat and a wavelet-based signal processing method. Pb(II) and Cu(II) concentrations were found to be in the range 3-7 µg L-1. Relative standard deviations were on average smaller than 3%, which is satisfactory in view of the trace concentration. For comparison, the determination was also carried out by using a high-end commercial equipment (μAutolab Type II, Eco Chemie). The average and standard deviation of the results were similar to those obtained with the lab-made potentiostat. No significant statistical difference was verified according to a paired t-test at a confidence level of 95%.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130106016ZK.pdf 288KB PDF download
  文献评价指标  
  下载次数:10次 浏览次数:3次