期刊论文详细信息
Journal of the Brazilian Chemical Society
A biosensor based on immobilization of lactate oxidase in a PB-CTAB film for FIA determination of lactate in beer samples
Denise Lowinsohn1  Mauro Bertotti1 
[1] ,Universidade de São Paulo Instituto de Química São Paulo SP ,Brazil
关键词: lactate;    biosensor;    prussian blue modified electrodes;    CTAB;    FIA amperometric detection;    beer;   
DOI  :  10.1590/S0103-50532008000400005
来源: SciELO
PDF
【 摘 要 】

An amperometric lactate biosensor with lactate oxidase immobilized into a Prussian Blue (PB) modified electrode was fabricated. The advantage of using cetyltrimethylammonium bromide (CTAB) in the electrodeposition step of PB films onto glassy carbon surfaces was confirmed taking into account both the stability and sensitivity of the measurements. The biosensor was used in the development of a FIA amperometric method for the determination of lactate. Under optimal operating conditions (pH = 6.9, E = -0.1 V), the linear response of the method was extended up to 0.28 µmol L-1 lactate with a limit of detection of 0.84 mmol L-1. The repeatability of the method for injections of a 0.28 mmol L-1 lactate solution was 2.2 % (n = 18). The usefulness of the method was demonstrated by determining lactate in beer samples and the results were in good agreement with those obtained by using a reference spectrophotometric enzyme method.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130105665ZK.pdf 276KB PDF download
  文献评价指标  
  下载次数:8次 浏览次数:5次