期刊论文详细信息
Journal of the Brazilian Chemical Society
Determination of rhodium and platinum by electrothermal atomic absorption spectrometry after preconcentration with a chelating resin
F. Sánchez Rojas1  C. Bosch Ojeda1  J.m. Cano Pavón1 
[1] ,University of Málaga Faculty of Sciences Department of Analytical ChemistryMálaga,Spain
关键词: rhodium;    platinum;    electrothermal atomic absorption spectrometry;    preconcentration;    flow injection;   
DOI  :  10.1590/S0103-50532007000600024
来源: SciELO
PDF
【 摘 要 】

The present publication describes a separation/preconcentration procedure for the subsequent trace analysis of rhodium and platinum in environmental samples. A chelating ion-exchange resin was employed for this purpose and an automatic on-line flow injection electrothermal atomic absorption spectrometry (FI-ET-AAS) method was used for the determination of trace amounts of Rh and Pt in different materials. The solutions are simultaneously treated and the ions determined sequentially. Detection limits of rhodium and platinum are 0.8 and 1.0 ng mL-1, respectively. Linearity was obtained in the range 0-50 ng mL-1 for rhodium and 0-100 ng mL-1 for platinum. The relative standard deviations were 1.8% for 10 ng mL-1 of Rh and 1.6% for 10 ng mL-1 of Pt. Enrichment factors were 20 for Rh and 14 for Pt.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130105530ZK.pdf 140KB PDF download
  文献评价指标  
  下载次数:9次 浏览次数:9次