期刊论文详细信息
Journal of the Brazilian Chemical Society
Colorimetric determination of ambient ozone using indigo blue droplet
Erika P. Felix1  Arnaldo A. Cardoso1 
[1] ,Universidade Estadual Paulista Instituto de Química Araraquara SP ,Brazil
关键词: ozone;    colorimetric sensor;    optical fibers;    liquid droplet;   
DOI  :  10.1590/S0103-50532006000200012
来源: SciELO
PDF
【 摘 要 】

A simple and sensitive method based on a liquid droplet is described for the measurement of atmospheric ozone. A 30 µL drop of indigo blue solution is suspended in a flowing-air sampling stream. The ozone collected reacts with the indigo solution resulting in its decolorization. The colorimetric sensor is composed of two optical fibers and the source of monochromatic light was a red LED (625 nm). The calibration curve was constructed with ozone standard concentrations ranging from 37 - 123 ppbv. The detection limit achieved was 7.3 ppbv. The method considered here showed itself to be easy to apply with a fast response and a total analysis time of only 5 minutes.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130105170ZK.pdf 177KB PDF download
  文献评价指标  
  下载次数:11次 浏览次数:14次