期刊论文详细信息
Journal of the Brazilian Chemical Society
Developing a continuous flow-square wave voltammetry method for determination of atrazine in soil solutions using the hanging mercury drop electrode
Luciana B. O. Dos Santos1  Gilberto Abate1  Jorge C. Masini1 
[1] ,Universidade de São Paulo Instituto de Química São Paulo SP ,Brazil
关键词: continuous flow;    atrazine;    soil solution;    square wave voltametry;   
DOI  :  10.1590/S0103-50532006000100005
来源: SciELO
PDF
【 摘 要 】

This work describes the development of a Continuous Flow-Square Wave Voltammetry method for determination of atrazine using the hanging mercury drop electrode. The best signal to noise ratio was obtained at the square wave frequency of 350 Hz and flow rate of 0.47 mL min-1. Under these conditions, the analytical curve obtained in 0.010 mol L-1 CaCl2 soil extracts in presence of 40 mmol L-1 BR buffer and 0.25 mol L-1 NaNO3 was linear for atrazine concentrations between 0.10 and 2.0 µg mL-1, with detection and quantification limits of 0.030 and 0.10 µg mL-1, respectively. The proposed method increased the analytical throughput in comparison with the batch methodology, allowing a sampling frequency of 72 h-1 to be accomplished. Besides, the sample consumption is significantly reduced, and only 341 µL are necessary for each analysis. The results obtained were similar to the ones obtained by HPLC, but the proposed method is faster and does not use organic solvents.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130105133ZK.pdf 172KB PDF download
  文献评价指标  
  下载次数:1次 浏览次数:8次