Laser Doppler anemometry measurements of steady flow through two bi-leaflet prosthetic heart valves | |
Ovandir Bazan2  Jayme Pinto Ortiz2  Francisco Ubaldo Vieira Junior1  Reinaldo Wilson Vieira1  Nilson Antunes1  Fabio Bittencourt Dutra Tabacow2  Eduardo Tavares Costa1  Orlando Petrucci Junior1  | |
[1] ,Universidade de São Paulo Polytechnic School Mechanical Engineering DepartmentSão Paulo SP ,Brazil | |
关键词: Heart valve prosthesis; Blood flow velocity; Laser-Doppler flowmetry; Próteses valvulares cardíacas; Velocidade do fluxo sanguíneo; Fluxometria por laser-Doppler; | |
DOI : 10.5935/1678-9741.20130076 | |
来源: SciELO | |
【 摘 要 】
INTRODUCTION: In vitro hydrodynamic characterization of prosthetic heart valves provides important information regarding their operation, especially if performed by noninvasive techniques of anemometry. Once velocity profiles for each valve are provided, it is possible to compare them in terms of hydrodynamic performance. In this first experimental study using laser doppler anemometry with mechanical valves, the simulations were performed at a steady flow workbench. OBJECTIVE: To compare unidimensional velocity profiles at the central plane of two bi-leaflet aortic prosthesis from St. Jude (AGN 21 - 751 and 21 AJ - 501 models) exposed to a steady flow regime, on four distinct sections, three downstream and one upstream. METHODS: To provide similar conditions for the flow through each prosthesis by a steady flow workbench (water, flow rate of 17L/min.) and, for the same sections and sweeps, to obtain the velocity profiles of each heart valve by unidimensional measurements. RESULTS: It was found that higher velocities correspond to the prosthesis with smaller inner diameter and instabilities of flow are larger as the section of interest is closer to the valve. Regions of recirculation, stagnation of flow, low pressure, and flow peak velocities were also found. CONCLUSIONS: Considering the hydrodynamic aspect and for every section measured, it could be concluded that the prosthesis model AGN 21 - 751 (RegentTM) is superior to the 21 AJ - 501 model (Master Series). Based on the results, future studies can choose to focus on specific regions of the these valves.
【 授权许可】
CC BY
All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202005130099299ZK.pdf | 2162KB | download |