Química Nova | |
Sistema multicomutado de análise em fluxo para determinação de amônio e monocloramina em águas residuais e produtos de desinfecção | |
Sivanildo Da Silva Borges2  Maria Soledad Moura Soares Fernandez Acevedo1  Mauro Korn1  | |
[1] ,Universidade Federal do Recôncavo da Bahia Centro de Ciências Exatas e Tecnológicas Cruz das Almas BA ,Brasil | |
关键词: monochloramine determination; ammonium determination; multicommutation; | |
DOI : 10.1590/S0100-40422009000500017 | |
来源: SciELO | |
【 摘 要 】
A binary sampling flow analysis system equipped with gas diffusion cell was developed for NH4+ and/or NH2Cl determination in wastewater and disinfection products samples based on the Berthelot reaction of the NH2Cl diffused through the semi-permeable PTFE membrane. The effect of the analytical conditions related to the reaction and flow parameters were evaluated and N-NH4+ and N-NH2Cl were determined in concentration ranges from 0.17 to 5 mg L-1 and from 0.5 to 14.5 mg L-1, respectively. Limits of detection (3σ) of 50 and 140 µg L-1 for N-NH4+ and N-NH2Cl were calculated, respectively, and RSD of 5 and 2% were calculated for 10 consecutive determinations of N-NH4+ (1 and 3 mg L-1) and N-NH2Cl (3 and 9 mg L-1), respectively with 30 determinations h-1.
【 授权许可】
CC BY-NC
All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202005130062825ZK.pdf | 191KB | download |