Rebound tonometry versus Goldmann tonometry in school children: feasibility and agreement of intraocular pressure measurements
Bruno Leonardo Barranco Esporcatte1  Flávio Siqueira Santos Lopes1  Camila Fonseca Netto1  Vespasiano Rebouças-santos1  Diego Torres Dias1  Fábio Iglesias Marujo1  Christiane Rolim-de-moura1 
关键词: Glaucoma/diagnosis;    Tonometry;    ocular;    Ocular hypertension;    Intraocular pressure;    Rebound effect;    Visual acuity;    Glaucoma/diagnóstico;    Tonometria ocular;    Hipertensão ocular;    Pressão intraocular;    Efeito rebote;    Acuidade visual;   
DOI  :  10.5935/0004-2749.20150095
来源: SciELO
PDF
【 摘 要 】

ABSTRACTPurpose: High intraocular pressure (IOP) is an important risk factor for a variety of pediatric ophthalmic conditions. The purpose of this study is to evaluate the feasibility, length of examination, and corneal epithelial damage induced by rebound tonometry (RBT) versus Goldmann applanation tonometry (GAT) in school children.Methods: Healthy children (n=57) participated in a randomized, transversal study with IOP measurement by GAT followed by RBT (study arm 1) or RBT followed by GAT (study arm 2). The number of attempts to acquire a reliable IOP measurement and the length of the examination were quantified. Corneal epithelial damage induced by tonometry was evaluated. Bland-Altman analysis was performed to establish the level of agreement between the two techniques.Results: The IOP was measured in all children with at least one of the devices. In both study arms, more children failed to be examined with GAT than with RBT (26% vs. 4%, and 16% vs. 6%, p<0.001, in study arm 1 and 2, respectively). The length of examination was shorter for RBT than for GAT (67.81 s ± 35.20 s vs. 126.70 s ± 56.60 s; p<0.0001); IOP measurements with RBT in both study arms were higher than those with GAT (15.20 ± 2.74 mmHg vs. 13.25 ± 2.47 mmHg, p=0.0247 and 16.76 ± 3.99 mmHg vs. 13.92 ± 2.08 mmHg, p=0.003, respectively). No difference was observed between RBT and GAT regarding the corneal epithelial damage caused by tonometry.Conclusion: IOP measurement is feasible in a greater number of children with RBT, and the examination was faster than that for GAT. Compared with GAT, RBT tended to overestimate the IOP. None of the methods induced marked corneal epithelial defects.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130003798ZK.pdf 495KB PDF download
  文献评价指标  
  下载次数:20次 浏览次数:16次