Evaluation of laparoscopic forceps jaw contact pressure and distribution using pressure sensitive film
Ikuo Yamamoto1  Murray John Lawn1  Takeshi Nagayasu2  Keitaro Matsumoto2  Maxime Maréchal3  Rui Zhu4 
[1] Engineering Department, Nagasaki University, Nagasaki, Japan;Graduate School of Biomedical Sciences, Nagasaki University, Nagasaki, Japa;Graduated School of Engineering, Nagasaki University, Nagasaki, Japan;Medical-Engineering Hybrid Professional Development Program, Graduate School of Biomedical Sciences, Nagasaki University, Nagasaki, Japan;
关键词: Contact pressure distribution;    tactile information;    pinch force;    grip force;   
DOI  :  10.1080/24699322.2019.1649073
来源: publisher
PDF
【 摘 要 】

In this study, the authors used the Fujifilm Prescale Pressure Measuring System to measure the contact pressure and distribution at the jaws of laparoscopic grasping forceps. This data was then correlated with measured pressures at the forceps handles to understand the relationship between the surgeon’s actuating pressure and that on the organ being manipulated. The purpose of this study is to create a database of tactile information to provide guidelines in defining minimally invasive surgery (MIS). This is expected to be important as today's society continues to progress in the use of automation, IoT, AI and MIS. In order to achieve the above, the authors developed an experimental device consisting of an actuator, a load cell and an MCU to stably actuate and control the handle side of grasping forceps. Target organs were simulated using triangular prisms of various silicone rubber materials. The experimental method involved actuating the handle side with preset pressure values for fixed time periods and using sensitive film to measure the pressure at the forceps tip. The film data was then scanned, processed and analyzed.

【 授权许可】

CC BY   

【 预 览 】
附件列表
Files Size Format View
RO202004239696260ZK.pdf 4528KB PDF download
  文献评价指标  
  下载次数:20次 浏览次数:19次