Sensors | |
Tetraazacyclohexadeca Macrocyclic Ligand as a Neutral Carrier in a Cr Ion-selective Electrode | |
Ashok Kumar Singh1  Rupam Singh1  | |
[1] Department of Chemistry, Indian Institute of Technology-Roorkee, Roorkee - 247 667, India | |
关键词: chromium-selective electrode; macrocyclic ionophore; potentiometric sensor; | |
DOI : 10.3390/s41200187 | |
来源: mdpi | |
【 摘 要 】
A polystyrene-based membrane of 2,10-dimethyl-4,12-diphenyl-1,5,9,13-tetraazacyclohexadeca-1,4,9,12-tetraene macrocyclic ionophore was prepared and investigated as Cr(III)-selective electrode. The best performance was observed with the membrane having the polystyrene-ligand-dibutylphthalate-sodiumtetraphenyl borate composition 1:4:1:1 with a Nernstian slope of 19.0 mV per decade of concentration between pH 3.0 and 6.5. This electrode has been found to be chemically inert and of adequate stability with a response time of 20 s and was used over a period of 100 d with good reproducibility (S= 0.3 mV). The membrane works satisfactorily in a partially non-aqueous medium up to a maximum 30% (v/v) content of methanol and ethanol. The potentiometric selectivity coefficient values indicate that the membrane sensor is highly selective for Cr(III) ions over a number of monovalent, divalent and trivalent cations. The membrane electrode has also been successfully used to determine Cr3+ in various food materials.
【 授权许可】
CC BY
© 2004 by the authors; licensee MDPI, Basel, Switzerland.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202003190060094ZK.pdf | 210KB | download |