期刊论文详细信息
Sensors
Electrochemical Oxidation and Determination of Oxalic Acid at an Exfoliated Graphite-Polystyrene Composite Electrode
Florica Manea3  Ciprian Radovan1  Ioana Corb3  Aniela Pop3  Georgeta Burtica3  Plamen Malchev2  Stephen Picken2 
[1] West University of Timisoara, Str.Pestalozzi, nr.16, 300115, Romania E-mail:;Delft University of Technology, 2600 GA Delft, The Netherlands; E-mail:;”Politehnica” University of Timisoara, P-ta Victoriei, nr.2, 300006, Timisoara, Romania; E-mail:
关键词: exfoliated graphite-polystyrene composite electrode;    oxalic acid;    electrochemical oxidation;    electrochemical determination;   
DOI  :  10.3390/s7040615
来源: mdpi
PDF
【 摘 要 】

An exfoliated graphite-polystyrene composite electrode was evaluated as an alternative electrode in the oxidation and the determination of oxalic acid in 0.1 M Na2SO4 supporting electrolyte. Using CV, LSV, CA procedures, linear dependences I vs. C were obtained in the concentrations range of oxalic acid between 0.5 to 3 mM, with LOD =0.05 mM, and recovery degree of 98%, without need of surface renewing between successive runs. The accuracy of the methods was evaluated as excellent comparing the detection results with that obtained using conventional KMnO4 titration method. In addition, the apparent diffusion coefficient of oxalic acid D was found to be around 2.89 · 10-8 cm2·s-1 by CA and CV.

【 授权许可】

Unknown   
© 2007 by MDPI (http://www.mdpi.org).

【 预 览 】
附件列表
Files Size Format View
RO202003190059406ZK.pdf 1864KB PDF download
  文献评价指标  