| Sensors | |
| On-Chip Immunoassay for Determination of Urinary Albumin | |
| Wanida Laiwattanapaisal1  Temsiri Songjaroen2  Thitima Maturos3  Tanom Lomas3  Assawapong Sappat3  | |
| [1] Centre for Excellence in Omics-Nano Medical Technology Project Development, Department of Clinical Chemistry, Faculty of Allied Health Sciences, Chulalongkorn University, Bangkok, 10330, Thailand;Graduate Programme in Clinical Biochemistry and Molecular Medicine, Faculty of Allied Health Sciences, Chulalongkorn University, Bangkok, 10330, Thailand; E-Mail:;Nanoelectronics and MEMS Laboratory, National Electronics and Computer Technology Centre, Pathum Thani, 12120, Thailand; E-Mails: | |
| 关键词: microalbuminuria; microfluidic system; on-chip immunoassay; poly(dimethylsiloxane); | |
| DOI : 10.3390/s91210066 | |
| 来源: mdpi | |
PDF
|
|
【 摘 要 】
An immunoassay performed on a portable microfluidic device was evaluated for the determination of urinary albumin. An increase in absorbance at 500 nm resulting from immunoagglutination was monitored directly on the poly(dimethylsiloxane) (PDMS) microchip using a portable miniature fibre-optic spectrometer. A calibration curve was linear up to 10 mg L–1 (r2 = 0.993), with a detection limit of 0.81 mg L–1 (S/N = 3). The proposed system showed good precision, with relative standard deviations (RSDs) of 5.1%, when evaluated with 10 mg L–1 albumin (
【 授权许可】
CC BY
© 2009 by the authors; licensee Molecular Diversity Preservation International, Basel, Switzerland.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO202003190055659ZK.pdf | 543KB |
PDF