期刊论文详细信息
Sensors
An Efficient and Compact Difference-Frequency-Generation Spectrometer and Its Application to 12CH3D/12CH4 Isotope Ratio Measurements
Kiyoshi Tsuji2  Hiroaki Teshima1  Hiroyuki Sasada1 
[1] SENTAN, Japan Science and Technology Agency/Sanbancho 5, Chiyoda-ku, Tokyo 102-0075, Japan;Department of Environmental Science and Technology, Interdisciplinary Graduate School of Science and Engineering, Tokyo Institute of Technology/4259, Nagatsuta-cho, Midori-ku, Yokohama, 226-8502, Japan; E-Mail:
关键词: isotope ratio;    methane;    laser spectroscopy;    difference frequency generation;    waveguide PPLN;   
DOI  :  10.3390/s100706612
来源: mdpi
PDF
【 摘 要 】

We have developed an efficient and compact 3.4 μm difference-frequency-generation spectrometer using a 1.55 μm distributed feedback (DFB) laser diode, a 1.06 μm DFB laser diode, and a ridge-waveguide periodically poled lithium niobate. It is continuously tunable in the 30 cm−1 span and is applied to 12CH3D/12CH4 isotope ratio measurements. The suitable pair of 12CH3D ν4 pP(7,6) and 12CH4 ν 24 R(6) F1(1) lines enabled us to determine their isotope ratio with a precision repeatability of 0.8‰ using a sample and a working standard of pure methane with an effective signal averaging time of 100 ms.

【 授权许可】

CC BY   
© 2010 by the authors licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190053134ZK.pdf 313KB PDF download
  文献评价指标  
  下载次数:11次 浏览次数:9次