Sensors | |
An Efficient and Compact Difference-Frequency-Generation Spectrometer and Its Application to 12CH3D/12CH4 Isotope Ratio Measurements | |
Kiyoshi Tsuji2  Hiroaki Teshima1  Hiroyuki Sasada1  | |
[1] SENTAN, Japan Science and Technology Agency/Sanbancho 5, Chiyoda-ku, Tokyo 102-0075, Japan;Department of Environmental Science and Technology, Interdisciplinary Graduate School of Science and Engineering, Tokyo Institute of Technology/4259, Nagatsuta-cho, Midori-ku, Yokohama, 226-8502, Japan; E-Mail: | |
关键词: isotope ratio; methane; laser spectroscopy; difference frequency generation; waveguide PPLN; | |
DOI : 10.3390/s100706612 | |
来源: mdpi | |
【 摘 要 】
We have developed an efficient and compact 3.4 μm difference-frequency-generation spectrometer using a 1.55 μm distributed feedback (DFB) laser diode, a 1.06 μm DFB laser diode, and a ridge-waveguide periodically poled lithium niobate. It is continuously tunable in the 30 cm−1 span and is applied to 12CH3D/12CH4 isotope ratio measurements. The suitable pair of 12CH3D ν4 pP(7,6) and 12CH4 ν 2+ν4 R(6) F1(1) lines enabled us to determine their isotope ratio with a precision repeatability of 0.8‰ using a sample and a working standard of pure methane with an effective signal averaging time of 100 ms.
【 授权许可】
CC BY
© 2010 by the authors licensee MDPI, Basel, Switzerland.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202003190053134ZK.pdf | 313KB | download |