Hemocompatibility of Inorganic Physical Vapor Deposition (PVD) Coatings on Thermoplastic Polyurethane Polymers
Juergen M. Lackner3  Wolfgang Waldhauser3  Paul Hartmann3  Franz Bruckert2  Marianne Weidenhaupt2  Roman Major4  Marek Sanak5  Martin Wiesinger1 
[1] Faculty of Technical and Environmental Sciences, University of Applied Sciences Upper Austria, Franz-Fritsch-Straße 11, Wels 4600, Austria;Grenoble Institute of Technology, Parvis Louis Néel, 38000 Grenoble Cedex 9, France;Joanneum Research Forschungsges.m.b.H., Institute for Surface Technologies and Photonics, Functional Surfaces, Leobner Straße 94, Niklasdorf A-8712, Austria;Institute of Metallurgy and Materials Sciences, Polish Academy of Sciences, Reymonta 25, Krakow 30-059, Poland;Department of Medicine, Jagiellonian University Medical College, Kopernika 23, Kraków 31-501, Poland;
关键词: inorganic thin film materials;    diamond-like carbon;    dynamic blood testing;    blood;   
DOI  :  10.3390/jfb3020283
来源: mdpi
PDF
【 摘 要 】

Biocompatibility improvements for blood contacting materials are of increasing interest for implanted devices and interventional tools. The current study focuses on inorganic (titanium, titanium nitride, titanium oxide) as well as diamond-like carbon (DLC) coating materials on polymer surfaces (thermoplastic polyurethane), deposited by magnetron sputtering und pulsed laser deposition at room temperature. DLC was used pure (a-C:H) as well as doped with silicon, titanium, and nitrogen + titanium (a-C:H:Si, a-C:H:Ti, a-C:H:N:Ti). In-vitro testing of the hemocompatibility requires mandatory dynamic test conditions to simulate in-vivo conditions, e.g., realized by a cone-and-plate analyzer. In such tests, titanium- and nitrogen-doped DLC and titanium nitride were found to be optimally anti-thrombotic and better than state-of-the-art polyurethane polymers. This is mainly due to the low tendency to platelet microparticle formation, a high content of remaining platelets in the whole blood after testing and low concentration of platelet activation and aggregation markers. Comparing this result to shear-flow induced cell motility tests with e.g., Dictostelium discoideum cell model organism reveals similar tendencies for the investigated materials.

【 授权许可】

CC BY   
© 2012 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190045256ZK.pdf 902KB PDF download
  文献评价指标  
  下载次数:16次 浏览次数:17次