期刊论文详细信息
Journal of Functional Biomaterials
Characterization of Porous TiO2 Surfaces Formed on 316L Stainless Steel by Plasma Electrolytic Oxidation for Stent Applications
Zhiguang Huan2  Lidy E. Fratila-Apachitei1  Iulian Apachitei2 
[1] Department of BioMechanical Engineering, Delft University of Technology, Mekelweg 2, Delft 2628 CD, The Netherlands;
关键词: drug eluting stent;    plasma electrolytic oxidation;    titanium oxide layer;    stainless steel;    surface porosity;   
DOI  :  10.3390/jfb3020349
来源: mdpi
PDF
【 摘 要 】

In this study, a porous oxide layer was formed on the surface of 316L stainless steel (SS) by combining Ti magnetron sputtering and plasma electrolytic oxidation (PEO) with the aim to produce a polymer-free drug carrier for drug eluting stent (DES) applications. The oxidation was performed galvanostatically in Na3PO4 electrolyte. The surface porosity, average pore size and roughness varied with PEO treatment duration, and under optimum conditions, the surface showed a porosity of 7.43%, an average pore size of 0.44 µm and a roughness (Ra) of 0.34 µm. The EDS analyses revealed that the porous layer consisted of Ti, O and P. The cross-sectional morphology evidenced a double-layer structure, with a porous titania surface and an un-oxidized dense Ti film towards the interface with 316L SS. After the PEO treatment, wettability and surface free energy increased significantly. The results of the present study confirm the feasibility of forming a porous TiO2 layer on stainless steel by combining sputtering technology and PEO. Further, the resultant porous oxide layer has the potential to be used as a drug carrier for DES, thus avoiding the complications associated with the polymer based carriers.

【 授权许可】

CC BY   
© 2012 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190044087ZK.pdf 965KB PDF download
  文献评价指标  