Use of a Combined SpO2/PtcCO2 Sensor in the Delivery Room
Serena Antonia Rubortone2  Maria Pia De Carolis1  Serafina Lacerenza2  Iliana Bersani2  Federica Occhipinti2 
[1] Division of Neonatology, Department of Paediatrics, Catholic University of Sacred Heart, Largo Gemelli 8, I-00168 Rome, Italy;
关键词: oxygen saturation;    partial pressure of carbon dioxide;    pulse oximeter;    delivery room;    TOSCA sensor;    neonate;   
DOI  :  10.3390/s120810980
来源: mdpi
PDF
【 摘 要 】

Arterial oxygen saturation (SaO2) and partial arterial pressure of carbon dioxide (PaCO2) are important respiratory parameters in critically ill neonates. A sensor combining a pulse oximeter with the Stow-Severinghaus electrode, required for the measurement of peripheral oxygen saturation (SpO2) and transcutaneous partial pressure of carbon dioxide (PtcCO2), respectively, has been recently used in neonatal clinical practice (TOSCA500ÒRadiometer). We evaluated TOSCA usability and reliability in the delivery room (DR), throughout three different periods, on term, late-preterm, and preterm neonates. During the first period (period A), 30 healthy term neonates were simultaneously monitored with both TOSCA and a MASIMO pulse oximeter. During the second period (period B), 10 healthy late-preterm neonates were monitored with both TOSCA and a transcutaneous device measuring PtcCO2 (TINAÒ TCM3, Radiometer). During the third period (period C), 15 preterm neonates were monitored with TOSCA and MASIMO after birth, during stabilization, and during transport to the neonatal intensive care unit (NICU). Blood gas analyses were performed to compare transcutaneous and blood gas values. TOSCA resulted easily and safely usable in the DR, allowing reliable noninvasive SaO2 estimation. Since PtcCO2 measurements with TOSCA required at least 10 min to be stable and reliable, this parameter was not useful during the early resuscitation immediately after birth. Moreover, PtcCO2 levels were less precise if compared to the conventional transcutaneous monitoring. However, PtcCO2 measurement by TOSCA was useful as trend-monitoring after stabilization and during transport to NICU.

【 授权许可】

CC BY   
© 2012 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190042477ZK.pdf 321KB PDF download
  文献评价指标  
  下载次数:27次 浏览次数:14次