期刊论文详细信息
Sensors
Performance Improvement of the One-Dot Lateral Flow Immunoassay for Aflatoxin B1 by Using a Smartphone-Based Reading System
Sangdae Lee2  Giyoung Kim1 
[1] Post-Harvest & Food Engineering Division, Department of Agricultural Engineering, National Academy of Agricultural Sciences, Suwon 441707, Korea;
关键词: lateral flow immunoassay;    aflatoxin B1;    Smartphone;    point-of-care;   
DOI  :  10.3390/s130405109
来源: mdpi
PDF
【 摘 要 】

This study was conducted to develop a simple, rapid, and accurate lateral flow immunoassay (LFIA) detection method for point-of-care diagnosis. The one-dot LFIA for aflatoxin B1 (AFB1) was based on the modified competitive binding format using competition between AFB1 and colloidal gold-AFB1-BSA conjugate for antibody binding sites in the test zone. A Smartphone-based reading system consisting of a Samsung Galaxy S2 Smartphone, a LFIA reader, and a Smartphone application for the image acquisition and data analysis. The detection limit of one-dot LFIA for AFB1 is 5 μg/kg. This method provided semi-quantitative analysis of AFB1 samples in the range of 5 to 1,000 μg/kg. Using combination of the one-dot LFIA and the Smartphone-based reading system, it is possible to conduct a more fast and accurate point-of-care diagnosis.