期刊论文详细信息
Sensors
Static Hyperspectral Fluorescence Imaging of Viscous Materials Based on a Linear Variable Filter Spectrometer
Patrik J. Murr1  Michael Schardt2 
[1] Institute for Measurement Systems and Sensor Technology, Technische Universität München, Theresienstrasse 90/N5, Munich 80333, Germany;
关键词: fluorescence;    hyperspectral imaging;    linear variable filter;    CMOS 2D sensor array;   
DOI  :  10.3390/s130912687
来源: mdpi
PDF
【 摘 要 】

This paper presents a low-cost hyperspectral measurement setup in a new application based on fluorescence detection in the visible (Vis) wavelength range. The aim of the setup is to take hyperspectral fluorescence images of viscous materials. Based on these images, fluorescent and non-fluorescent impurities in the viscous materials can be detected. For the illumination of the measurement object, a narrow-band high-power light-emitting diode (LED) with a center wavelength of 370 nm was used. The low-cost acquisition unit for the imaging consists of a linear variable filter (LVF) and a complementary metal oxide semiconductor (CMOS) 2D sensor array. The translucent wavelength range of the LVF is from 400 nm to 700 nm. For the confirmation of the concept, static measurements of fluorescent viscous materials with a non-fluorescent impurity have been performed and analyzed. With the presented setup, measurement surfaces in the micrometer range can be provided. The measureable minimum particle size of the impurities is in the nanometer range. The recording rate for the measurements depends on the exposure time of the used CMOS 2D sensor array and has been found to be in the microsecond range.

【 授权许可】

CC BY   
© 2013 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190032890ZK.pdf 278KB PDF download
  文献评价指标  
  下载次数:9次 浏览次数:1次