期刊论文详细信息
Sensors
Programmable Gain Amplifiers with DC Suppression and Low Output Offset for Bioelectric Sensors
Albano Carrera2  Ramón de la Rosa1 
[1] Laboratory of Electronics and Bioengineering, ETSI de Telecomunicación, Universidad de Valladolid, Campus Miguel Delibes, Paseo Belén, 15. Valladolid 47011, Spain;
关键词: bioelectric sensors;    biomedical electronics;    amplifiers;    high-pass filters;   
DOI  :  10.3390/s131013123
来源: mdpi
PDF
【 摘 要 】

DC-offset and DC-suppression are key parameters in bioelectric amplifiers. However, specific DC analyses are not often explained. Several factors influence the DC-budget: the programmable gain, the programmable cut-off frequencies for high pass filtering and, the low cut-off values and the capacitor blocking issues involved. A new intermediate stage is proposed to address the DC problem entirely. Two implementations were tested. The stage is composed of a programmable gain amplifier (PGA) with DC-rejection and low output offset. Cut-off frequencies are selectable and values from 0.016 to 31.83 Hz were tested, and the capacitor deblocking is embedded in the design. Hence, this PGA delivers most of the required gain with constant low output offset, notwithstanding the gain or cut-off frequency selected.

【 授权许可】

CC BY   
© 2013 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190032818ZK.pdf 1208KB PDF download
  文献评价指标  
  下载次数:5次 浏览次数:11次